Medical pendants are
widely used in hospital because of its convenient operation, highly integrated
functions, high application range and excellent space utilization. As
professional medical pendant manufacturer, Mingtai company gives you a general
understanding of the use of medical pendants and solutions to common faults.
1. Arm, terminal box and
instrument platform are not flexible
Common reasons:
(1). The nut at the upper
joint bearing of the pipe is too tight.
(2). The nut at the middle
beam joint bearing is too tight.
Solution:
Adjust the tightening nut.
2. The rotation angle of the
arm, terminal box and instrument platform exceeds 340°, and cannot be limited.
Common reasons:
(1). The positioning pin is
broken becaue of excessive force.
(2). The positioning pin is
detached.
Solution:
Replace the positioning pin.
3. Drifting; medical pendants sometimes
appear to drift, which will cause unnecessary trouble to surgery, affecting the
efficiency of surgery.
Common reasons:
Usually because the base of
the pendant is not flat, or the bracket is shaken.
Solution:
Reinforce the pendant
bracket.
4. The button does not work,
the work indicator does not light, or the work indicator light, but the brake
and lift do not work.
Common reasons:
(1). Switching power supply
is damaged
(2). The overload protection
circuit work because the instrument platform is overloaded.
(3). The motor is
short-circuited to make the overload protection circuit work.
(4). Poor contact of the
connector or short circuit of the control line causes no control signal to the
input end of the electrical box and causes no action
(5). The board is damaged or
the components are in poor contact and the solder is soldered.
Solution:
(1). Replace the switching
power supply
(2). Remove the heavy objects
on the instrument platform
(3). Repair or replace the
motor
(4). Replace the connector,
control line
(5). Replace the board
Common reasons:
(1). The medical gas terminal
is not reset or the terminal seal is damaged.
(2). There is debris in the
gas source to catch the terminal.
Solution:
(1). Repeatedly plugging and
unplugging with a secondary terminal plug to reset it
(2). Replace the medical gas
terminal
(3). Remove the debris in the
gas terminal